Dolphin G3

Dolphin G3 Specifications

User-Friendly Operation

Dolphin G3™ guides the technician through all the steps necessary to carry out a job – tube sheet mapping, parameter input, measurements, verification and report generation. An intuitive graphic user interface, including helpful right-click functions, makes scrolling, zooming and navigation easy every step of the way.

- Mapping

The system creates a computerized model of the heat exchanger tube sheet based on a photo, CAD file or scanned schematic diagram. Automatic numbering and mapping is used to establish test plans and track inspection progress.functions, makes scrolling, zooming and navigation easy every step of the way.

- Report Generation

Dolphin G3™ presents an easy-to-read online report (with drill down options) to the technician for verification, as well as a final summary report including tabular reporting and screenshots of the detected flaws in PDF or HTML format.

- Logging

All measurements are logged in their raw form. Different thresholds and algorithms can be applied to regenerate reports with varying levels of sensitivity.

- One-click Collaboration

The system utilizes only a single calibration tube (supplied), eliminating inventory requirements and typical calibration challenges.

- Visual Aids For Detection

Programmable thresholds for different flaw types and sizes can be overlaid on the measured signals to assist in visual evaluation of defects.

Dolphin G3 Specifications

Tube Size 5/16” up to 4” (8mm – 102mm) inner diameter.
Tube Configuration Any configuration including U-bends, finned tubes, twisted tubes and spiral wound tubes.
Tube Material Any material including: metals (ferrous and non-ferrous) and non-metals (graphite,plastics, ceramics).
Maximum Length Up to 80’ length depending on defect size, tube configuration and type.
Detectable Defects
5/16”-2 1/2” (8mm – 63.5mm) Tubes 2 1/2”-4” (63.5mm – 102mm) Tubes
Holes
  • Minimum diameter 0.039” (1mm) Even smaller in many cases
  • Minimum diameter 0.118” (3mm) Even smaller in many cases
Blockages
  • Minimum of 5% of cross section
  • Minimum of 5% of cross section
Erosion
  • Minimum 5% of wall thickness
  • Minimum 5% of wall thickness
Pitting
  • Minimum 20% of wall thickness
  • Minimum 50% of wall thickness
Inspection Speed Under 10 seconds per tube depending on tube size, length and configuration.5 seconds per tube less than 20 ft (6m).
Hardware
  • Compact handheld device:

    Non-invasive probe assembly including a transducer, microphone, controls, LCD screen and adaptors to fit tubes of any internal diameter up to 4”. Customized extensions (optional) for easy access to tube openings in challenging situations.

  • Main measurement and control unit (MCU):

    External computer connected to handheld device. Utilized for inspection setup, probe status monitoring, test data recording and analysis via AcousticEye software.

Software Patented Acoustic Pulse Reflectometry (APR) technology featuring specialized, proprietary algorithms for NDT inspection.
Reporting Customizable, graphical on-line reports. Available output in PDF or HTML format.
Physical Characteristics Compact, rugged and lightweight design. Total box weight: 39.7 lbs (18Kg) Box dimensions: 22” L x 17.9” W x 10.4” H (56 x 45.5 x 26.5 cm)
Power Input Dual voltage system (110V/220V)
Temperature Range 14° to 122° F (-10° to +50° C)
Certifications CE Declaration of Conformity; Safety Certificate IEC 61010; EMC Test Certificate; Company Quality System certified to ISO 9001:2008

Affiliate Companies

Ready Oilfield
OMNI
ARISE

Accreditation